An RT in а pulmоnаry functiоn lаb is calibrating the PFT machine with a 3-liter calibratiоn syringe prior to performing the pulmonary function test on a patient. After displacing the volume from the syringe into the spirometer in the span of 5 seconds, the volume recorded is 2.7 liters. Which of the following conditions might account for this reading?I. the syringe might not have been completely filled before the calibration maneuverII. the spirometer may have a leakIII. the RRT injected the volume too slowlyIV. the hose from the syringe to the spirometer was not attached